Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 595
Technical data
Electromagnetic immunity
Immunity
against...
IEC 60601-1-2 test level Compliance
level (of device)
Electromagnetic environment
Electrostatic
discharge, ESD
(IEC 61000-4-2)
Contact discharge: ±6kV
Air discharge: ±8kV
±6kV
±8kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transients / bursts
(IEC 61000-4-4)
PS250 / P2500 lines:
±2kV
Longer input / output
lines: ± 1 kV
±2kV
±1kV
Mains power quality should be of a
typical commercial or hospital
environment.
Surges on AC
mains lines
(IEC 61000-4-5)
Common mode: ±2kV
Differential mode: ±1kV
±2kV
±1kV
Mains power quality should be a
typical commercial or hospital
environment.
Power frequency
magnetic field
50/60 Hz
(IEC 61000-4-8)
3 A/m 3 A/m Equipment emitting high levels of
power line magnetic fields (in excess
of 3A/m) should be kept at a distance
to reduce likelihood of interference.