Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 597
Technical data
Conducted RF
RF coupled into
lines
(IEC 61000-4-6)
Radiated RF
(IEC 61000-4-3)
150 kHz to 80 MHz
80 MHz to 2.5 GHz
3 Vrms
3V/m
[V
1] V
[E1] V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter as below.
Recommended separation distance
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
Electromagnetic immunity
Immunity
against...
IEC 60601-1-2 test level Compliance
level (of device)
Electromagnetic environment
d
3.5
E
1
-------
P=
80 MHz to 800 MHz
d
7
E
1
------
P=
800 MHz to 2.5 GHz