Do you have a question about the Dräger JM-105 and is the answer not in the manual?
Explains the Dräger Jaundice Meter JM-105 as a screening device for neonatal jaundice.
Details how the meter identifies neonates at risk for hyperbilirubinemia based on assessment.
Outlines the daily operational checkout procedure for the Jaundice Meter JM-105.
Describes how to configure device settings for memory and measurement averaging.
Provides step-by-step instructions for performing a TcB measurement.
Explains how to plot TcB results on a nomogram to determine percentile ranges.
Details the training requirements and process for personnel using the Jaundice Meter JM-105.
Recommends validating the device against serum bilirubin readings for accuracy.
The Dräger Jaundice Meter JM-105 is a screening device designed for the non-invasive assessment of jaundice in newborns. It provides a transcutaneous measurement of bilirubin (TcB) in mg/dL or µmol/L, helping to identify neonates who may require a serum bilirubin (TSB) measurement. This device is intended to be used as part of a systematic assessment of risk factors for hyperbilirubinemia, allowing for the identification of infants at increased risk for more severe hyperbilirubinemia in their first week of life, thereby prompting closer monitoring.
The primary function of the Dräger Jaundice Meter JM-105 is to measure transcutaneous bilirubin levels. It does this by emitting light into the skin and analyzing the reflected light to estimate the bilirubin concentration. This method offers a quick and non-invasive alternative to traditional blood draws for initial screening. The device is particularly useful for identifying neonates who are at risk for hyperbilirubinemia, a condition characterized by elevated bilirubin levels that can lead to jaundice. By providing immediate TcB readings, the JM-105 helps healthcare professionals make timely decisions regarding further diagnostic testing, such as serum bilirubin measurements. The device is designed to be user-friendly, with a clear display that shows the measured bilirubin values. It can be configured to display results in either mg/dL or µmol/L, accommodating different clinical standards.
The Dräger Jaundice Meter JM-105 offers several features to facilitate its use in various clinical settings.
Screening Protocols: Two main protocols are suggested for its use:
Measurement Site Selection: The device allows for flexibility in choosing the measurement site, depending on the clinical environment:
Measurement Procedure: Taking a transcutaneous bilirubin measurement involves a straightforward process:
Out-of-Range Indicators: If a measured value falls outside the device's measuring range (e.g., 0 to 20 mg/dL or 0 to 340 µmol/L), a blinking value will appear on the display. The specific blinking indicator (e.g., "-0-", ">20", or ">340") depends on the user's preference settings.
Nomogram Plotting: Results from the JM-105 can be plotted on a transcutaneous bilirubin nomogram, specifically designed for neonates up to 96 hours of age (≥35 weeks of gestation). This nomogram helps visualize the baby's TcB value in relation to percentile ranges, aiding in the assessment of risk for hyperbilirubinemia. For example, a TcB of 10 mg/dL at 50 hours of age might place a baby at the edge of the 95th percentile, indicating a need for a TSB measurement. Subsequent TcB measurements can also be plotted to detect trends, such as a rapidly rising TcB.
The Dräger Jaundice Meter JM-105 incorporates features to ensure its proper functioning and reliability, primarily through a daily operational checkout procedure.
Daily Operational Checkout: Before each use, a daily operational checkout procedure must be performed:
Configuration Options: The device allows for configuration of several settings:
Cleaning: The probe tip should be cleaned with an alcohol swab before taking a transcutaneous bilirubin measurement, contributing to hygiene and accurate readings.
Staff Competency Validation: To ensure proper use, all clinical personnel responsible for JM-105 testing must undergo training. This includes receiving a demonstration or viewing a training video, reading the Operating Instructions Manual, performing a return demonstration on infants, and completing a Skills Checklist. Successful completion is documented in the employee's education record.
New or Serviced Device Validation: It is recommended that users validate new or serviced JM-105 devices within their patient populations. This involves taking TcB readings on infants scheduled for TSB analysis, ensuring the time gap between TcB and blood sample collection is not greater than one hour. Recording these simultaneous readings helps develop a facility-specific protocol for the device's use.
| Type | Transcutaneous Jaundice Meter |
|---|---|
| Display | LCD |
| Data Storage | Up to 100 measurements |
| Power Supply | Rechargeable battery |
| Accuracy | ±1.5 mg/dL |
| Storage Temperature | -20°C to +50°C |












Loading...