Operation
In the Event of a Power Failure
Alarms
Operating Instructions Oxylog 2000, SW 3.n
37
In the Event of a Power Failure
Alarms
In the event of an alarm
1 The red alarm status will light up / will flash,
the audible alarm will sound, either continuously,
intermittently or as a single beep (every 30 seconds),
an alarm message will appear on the second line of
the backlit display.
Assess the audible alarm:
Continuous or intermittent tone = Urgent! Immediate
action is required!
Single beep (every 30 seconds) = Advisory message.
Read alarm message in display and correct condition
using the guidelines from chapter "Troubleshooting"
on pages 45, 46.
NOTE: The audible alarm may be silenced for 2 minutes
during an alarm:
2 Press »g Reset« key.
If the alarm situation persists, the audible alarm will
resume after 2 minutes – or immediately if a new,
higher priority alarm occurs.
WARNING !
Continuous or intermittent audible alarms
require immediate operator attention to avert
or to prevent development of situations with
the possibility of patient injury.
WARNING !
Automatic ventilation, volume measurement,
and alarms do not operate in the event of a
power failure!
Spontaneous breathing can continue through
the integrated demand valve.
Immediately start ventilating patient with
independent manual ventilation device
(resuscitation bag) using PEEP and/or
increased inspiratory oxygen concentration
where necessary and appropriate!