Technical Data
Performance Data
Operating Instructions Oxylog 2000, SW 3.n
52
Characteristics for support of spontaneous breathing
Demand valve pressure
trigger level approximately - 1 cmH2O
Max. demand flow (at -4 cmH2O) 100 L/min
Sensitivity of synchronization
for SIMV 4 L/min
for Assist/Control 3 L/min
Ventilator compliance
with 1.5 m (5 ft) patient circuit < 1.0 mL/cmH2O
with 3.0 m (10 ft) patient circuit < 1.2 mL/cmH2O
Inspiratory resistance < 6 cmH2O/L/s
Expiratory resistance < 4 cmH2O/L/s
Dead space including flow sensor approximately 28 mL
Measuring ranges
Pressure gauge - 10 to 80 cmH2O
Display accuracy ± 2 cmH2O
Flow measurement range 2 to 120 L/min bidirectional
Max. permissible differential
pressure for flow measurement ± 4 cmH2O
Resistance of flow sensor 3 cmH2O at 100 L/min
Measurement of minute ventilation
Range 1 to 40 L/min
Accuracy (with 100% O2 by vol.,
1013 hPa, 20ºC, 50% relative
humidity)
for 1 to 5 L/min ± 1 L/min
for 5 to 40 L/min greater of ± 12% of measured value or ± 1 L/min
Patient connector 22 mm male ISO tapered connector
O2 concentration of the
ventilation gas (with O2 supply)
Switch set to »Air Mix«
for MV less than 7 L/min O2 concentration can increase to 90 vol.%
for MV greater than 7 L/min 60 % O2 by vol. ±10 %
Switch set to »No Air Mix« 100 % O2 by vol.
Response in extreme conditions:
when supply pressure is 10 bar
(147 psi) The applied tidal volume VT increases additionally
by approximately 5% of the set value