Prepare the device for the entire
system test
• Press the power on button on the system
screen.
• Occlude the Y piece on the occlusion port
• Set the APL valve to 30.
• Close the Add O
2
.
• Confi rm the checklist if all the points are
checked and ok.
Entire system test
• Press the entire system test and confi rm by
pressing the rotary knob to start the test.
• The entire system test includes a pre-test.
This will identify any common errors that
occur during the self test (for example,
incorrect setting of the APL valve) within
60 seconds, allowing the user to rectify them
quickly before continuing the entire system
test. In the event of such a failure, resolve
the issue and repeat the test.
The leakage assistant will assist in
identifying the location of any leaks that
occur in the breathing system. Press the
leakage assistant button and follow the on
screen instructions.
• On completion of the pre-test, the entire test
system will continue automatically.
Leakage test
• Complete after changing breathing circuits
Breathing system test
• Complete after replacing the fl ow sensors
Start Up