Prepare the device for the entire
system test
Press the power on button on the system
screen.
Occlude the Y piece on the occlusion port
Set the APL valve to 30.
Close the Add O
2
.
Con rm the checklist if all the points are
checked and ok.
Entire system test
Press the entire system test and confi rm by
pressing the rotary knob to start the test.
The entire system test includes a pre-test.
This will identify any common errors that
occur during the self test (for example,
incorrect setting of the APL valve) within
60 seconds, allowing the user to rectify them
quickly before continuing the entire system
test. In the event of such a failure, resolve
the issue and repeat the test.
The leakage assistant will assist in
identifying the location of any leaks that
occur in the breathing system. Press the
leakage assistant button and follow the on
screen instructions.
On completion of the pre-test, the entire test
system will continue automatically.
Leakage test
Complete after changing breathing circuits
Breathing system test
Complete after replacing the fl ow sensors
Start Up