Start Up
Leakage assistant test
This should be carried out after the vaporisers
have been replaced or reusable soda lime
canister has been re lled. In addition, it should
be performed following the self test to check
the integrity of the breathing system including
the vaporisers.
On completion of the entire system test, the standby screen will appear
Potential test results
Test Result - Colour Test Result Message
Green Test successful Fully operable
Red Test unsuccessful Not operable
Yellow Test partially successful Not fully operable
Grey Not tested Not tested