Hypoxic guard function (SORC)
The Perseus A500 is fi tted with a hypoxic
guard function (SORC). This is an electronic
O
2
minimum delivery system that guarantees
at least 25% O
2
whenever N
2
O is used as the
carrier gas. This prevents the delivery of hypoxic
gas mixtures.
Selecting a ventilation mode
Spontaneous breathing (Spont)
Select man/spont and confi rm.
Adjust the APL valve to SPONT –
the breathing system is now open for
spontaneous breathing.
Manual ventilation (Man)
Select man/spont and confi rm
Switch the APL valve to MAN and adjust
accordingly.
Note - the breathing system pressure can
be quickly released by pulling the APL valve
upwards.
Mechanical ventilation
Select the ventilation mode required by
pressing the appropriate tab.
Set the ventilation parameters as required
by selecting the parameter, adjusting and
confi rming by pressing the rotary knob
Confi rm the ventilation mode by pressing
the rotary knob.
When changing between different modes,
the parameters from the previous mode are
transferred directly or suitably derived.
Operation

Other manuals for Dräger Perseus A500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Perseus A500 and is the answer not in the manual?

Dräger Perseus A500 Specifications

General IconGeneral
BrandDräger
ModelPerseus A500
CategoryMedical Equipment
LanguageEnglish

Related product manuals