EasyManuals Logo

Dräger RailGard-S1 User Manual

Dräger RailGard-S1
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
14
2. Adjusting the zero point
Apply zero gas (e.g. 99.99 % by Vol. N
2
) to the sensor, which corresponds
to ZERO (4 mA).
The output current can be checked with a voltmeter at the two test points
on the left hand side of the converter (“Measuring Point 4-20mA”).
Alternatively you may refer to the display of the Dräger RailGard, provided
that the corresponding channel configured correctly.
Use the potentiometer ("Adjust sensor zero point") to adjust the voltage at
the test point to 0.4 V (indication on the voltmeter).
The display of the Dräger RailGard shall indicate "0".
3. Adjusting the span
Apply span gas (e.g. 50 %LEL) to the sensor.
The output current can be checked with a voltmeter at the two test points
on the left hand side of the converter (“Measuring Point 4-20mA”).
Alternatively you may refer to the display of the Dräger RailGard, provided
that the corresponding channel configured correctly.
Use the potentiometer ("Adjust sensor sensitivity") to adjust the output
current to the value which corresponds to the span gas being used.
If a 50 %LEL reference gas is used, the display on the Dräger RailGard
shall indicates "50"; the voltmeter should indicate 1.2 V (corresponding to
12 mA).
Potentiometers and test points at the converter

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger RailGard-S1 and is the answer not in the manual?

Dräger RailGard-S1 Specifications

General IconGeneral
BrandDräger
ModelRailGard-S1
CategoryControl Unit
LanguageEnglish

Related product manuals