14
Operation
Calling the Info Mode
● In measuring mode, press the key for approx. 3 seconds.
Any information and fault codes, the peak values as well as the TWA
1)
and STEL
*)
exposure
values are displayed in succession (press the key for the next display). If any warning or
fault messages exist, the corresponding note or error codes are displayed (page 23 to
page 27).
— If no key is pressed for 10 seconds, the device returns automatically to measuring mode.
1) Only when activated in the device configuration. Delivery status: not activated.
Warning messages are displayed. Numerical codes of the warning messages: see
page 23.
⇓ -key
Fault messages are displayed. Numerical codes of the fault messages: see
page 25.
⇓ -key
The peak values = the maximum measured values in the case of e.g. CO, H
2
S, ...
or the minimum measured values in the case of O
2
within the storage interval are
displayed
⇓ key
The average values of the exposures based on a shift of e.g. 8 hours (TWA) of all
the active sensors for the exposure evaluation are displayed
⇓ key
The short-term values (STEL) = average values of the concentrations over the
average value duration of all the active sensors for the exposure evaluation are
displayed
⇓ key
The device is in measuring mode again
OK
OK