Maintenance
Calibrating / Adjusting the Device
Calibration may not be possible due to device and channel errors.
Allow the sensors to warm up before the calibration / adjustment!
Warm-up time: refer to the Instructions for Use of the installed DrägerSensors (on CD).
Calibration interval:
Observe the relevant specifications in the Instructions for Use of the installed DrägerSen-
sors.
For critical applications, observe the recommendations in EN 60079-29-2
1)
or
EN 45544-4
2)
and national regulations. We recommend that you calibrate all channels af-
ter 6 months.
Improve the zero point accuracy – carry out the fresh air calibration, page 36.
Set the sensitivity of all sensors to the value of the test gas – carry out the 1-button cali-
bration, page 38.
Set the sensor sensitivity to a value of the test gas – calibrate/adjust the sensitivity,
page 40.
1) EN 60079-29-2 – Gas measuring device - Selection, installation, use and maintenance of apparatus
for the measurement of combustible gases or oxygen.
2) EN 45544-4 – Electrical devices for the direct detection and direct concentration measurement of toxic
gases and vapours – Part 4: Guidelines for selection, installation, use and maintenance.
CAUTION
Do not inhale the test gas. Risk to health!
Observe the hazard warnings of the relevant Safety Data Sheets.

Table of Contents

Other manuals for Dräger X-am 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger X-am 2000 and is the answer not in the manual?

Dräger X-am 2000 Specifications

General IconGeneral
BrandDräger
ModelX-am 2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals