EasyManuals Logo

Dräger X-am 2000 Technical Manual

Dräger X-am 2000
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
31
Maintenance
The current gas concentration values and
the special symbol » « (for bump test)
flash.
Press the key to start the function test
with gas.
Open the test gas cylinder valve to let test
gas flow over the sensor.
If gas concentration exceeds the alarm
thresholds A 1 or A 2 the corresponding
alarm will occur.
Exit the function test with gas:
For the manual bump test:
After the set bump test concentration is
reached:
Press the key.
The display containing the current gas
concentration changes with the display
»O.
The bump test that was carried out is do-
cumented with the result and date in the
device memory.
Close the test gas cylinder valve and re-
move the device from the calibration cr-
adle.
If the concentration values have now fal-
len under the A1 alarm setpoints, the de-
vice returns to the measuring mode.
If during the function test no alarm occurs
within 1 minute and the gassing test was
not confirmed by pressing OK, the alarm
mode is activated to indicate failure.
The fault message » « appears and
» « is displayed instead of the
measured value on the faulty measuring
channel.
In this case, repeat the function test with
gas or calibrate the device, page 35.
02423854_01_en.eps
OK
02523854
_
01
_
en.e
p
s
OK
02623854_01_en.eps

Table of Contents

Other manuals for Dräger X-am 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger X-am 2000 and is the answer not in the manual?

Dräger X-am 2000 Specifications

General IconGeneral
BrandDräger
ModelX-am 2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals