31
Maintenance
The current gas concentration values and
the special symbol » « (for bump test)
flash.
Press the key to start the function test
with gas.
Open the test gas cylinder valve to let test
gas flow over the sensor.
If gas concentration exceeds the alarm
thresholds A 1 or A 2 the corresponding
alarm will occur.
Exit the function test with gas:
For the manual bump test:
After the set bump test concentration is
reached:
Press the key.
The display containing the current gas
concentration changes with the display
»O.
The bump test that was carried out is do-
cumented with the result and date in the
device memory.
Close the test gas cylinder valve and re-
move the device from the calibration cr-
adle.
If the concentration values have now fal-
len under the A1 alarm setpoints, the de-
vice returns to the measuring mode.
If during the function test no alarm occurs
within 1 minute and the gassing test was
not confirmed by pressing OK, the alarm
mode is activated to indicate failure.
The fault message » « appears and
» « is displayed instead of the
measured value on the faulty measuring
channel.
In this case, repeat the function test with
gas or calibrate the device, page 35.
02423854_01_en.eps
OK
02523854
_
01
_
en.e
p
s
OK
02623854_01_en.eps