44
Maintenance
Position the device on an even and level
surface.
Connecting the power pack to the mains.
1 The green "Mains" LED lights.
Insert the device into the charging modu-
le.
2 Display LED on the charger:
If a fault occurs:
Remove the device from the charging
module and insert it again.
If the fault still occurs, have the charging
module repaired.
It takes approx. 4 hours to fully charge an
empty rechargeable battery.
A short circuit of the charging contacts in the
charging modules, e.g. by metallic objects
that have fallen in, does not result in damage
to the charging station. It should, however,
be avoided due to possible heating hazards
and incorrect displays on the charging
module.
In the event of a short circuit or if the power
pack is overloaded:
3 The red "Overload" LED lights, and an
audible alarm sounds.
After the fault has been corrected, the
alarm is switched off automatically and
the charging process is restarted.
In the event of a power failure, the devi-
ces already charged will be protected
from discharging.
Charge
Fault
Full
00823854_01.eps
0
13 2

Table of Contents

Other manuals for Dräger X-am 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger X-am 2000 and is the answer not in the manual?

Dräger X-am 2000 Specifications

General IconGeneral
BrandDräger
ModelX-am 2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals