14 X-plore 8000
Transport
6 Transport
Transport in the original packaging or in optionally available
transport box.
7 Storage
Storing the whole system:
Remove respiratory filter and rechargeable battery.
Store the components in a container or cabinet dry and
clean and protect them from direct sunlight and thermal
radiation.
Storing rechargeable batteries:
Deeply discharged batteries may get damaged after prolonged
storage. Charge them to 50 to 70 % prior to storage.
If storage lasts for over 6 months, charge them in the
meantime.
Do not store rechargeable batteries for prolonged periods
outside the recommended temperature range. This might
reduce the remaining capacity and number of potential
charge cycles.
8 Disposal
Dispose of the device according to local waste disposal
regulations.
9 Technical data
9.1 Overall system
9.2 Rechargeable batteries
9.3 Standard battery charger
Flow rate of respiratory protective
device/helmet/visor
170/190/210 l/min
Flow rate of full facepiece 115/130/145 l/min
Rated period of service 4 hours
with standard
rechargeable battery
8 hours
with long-life
rechargeable battery
Operating temperature
1
14 °F to 140 °F (-10 °C
to 60 °C)
Working and storage humidity
1)
≤ 95 % relative humidity
Storage temperature
1)
-4 °F to 140 °F (-20 °C to
60 °C)
Noise approx. 64 dB(A)
International Protection Code IP 65
RFID
Technology Induction transfer
Frequency range 13553 to 13567 kHz
Broadcast transmission capacity -2,30 dBµA/m (10 m)
Bluetooth
Technology FHSS 2.4 GHz
(BT 2.1 + EDR)
Frequency range 2402.0 to 2483.5 MHz
Broadcast transmission capacity 0,97 mW / -0.14 dBm
EIRP
1 For the battery charger and rechargeable batteries, refer to the
separate information provided in this chapter.
For other components, refer to the corresponding Instructions for
Use.
Operating temperature 14 °F to 131 °F (-10 °C to
55 °C)
Operating/storage area humidity ≤ 95 % relative humidity
Storage temperature -4 °F to 122 °F (-20 °C to
50 °C)
Charging temperature 32 °F to 122 °F (0 °C to
50 °C)
Standard rechargeable battery
Charging time < 4 hours
Operational life time after a full
charge
approx. 4 hours
1)
Rated voltage 10,8 V
Rated capacity 3,35 Ah
Stored energy 36 Wh
Long-life rechargeable battery
Charging time < 4 hours
Operational life time after a full
charge
approx. 8 hours
1
1 Varies depending on the preset flow rate and the employed respi-
ratory filter and respiratory inlet covering type
Rated voltage 10,8 V
Rated capacity 6,70 Ah
Stored energy 72 Wh
Input voltage 15 V
Input current 4 A
Output voltage 9 - 12,6 V
Output current 4 A
International Protection Code IP 30
Operating temperature 32 °F to 122 °F (0 °C to
50 °C)