LT-2334
MAINTENANCE
IGO2 POC CARE AND CLEANING
WARNING
To avoid electric shock, do not remove the concentrator cover.
The cover should only be removed by a qualied Drive DeVilbiss
technician. Do not apply liquid directly to the cover or utilize any
petroleum-based solvents or cleaning agents.
Use of harsh chemicals (including alcohol) is not recommended.
If bactericidal cleaning is required, a non-alcohol based product
should be used to avoid inadvertent damage.
Use no lubricants, oils or grease.
Before attempting any cleaning procedures, turn the unit “Off” and
disconnect from AC or DC power.
Do not service or clean this device while in use with a Patient.
Exterior Cover & Battery
Clean the concentrator exterior cover and battery as needed:
1. Ensure battery is installed while cleaning the cover.
2. Use a damp cloth or sponge with a mild household cleaner on the
exterior cover, and wipe it dry. If battery is removed, wipe battery bay
with a Dry Cloth only.
CAUTION
Do not apply liquid directly to the cover.
3. Wipe battery with a Dry Cloth as needed.
Carrying Case
The carrying case should be cleaned as needed. To clean, follow these
steps:
1. Remove the device from the carrying case before cleaning.
2. As needed, clean the case by using a damp cloth with a mild
household cleaner and wipe dry.
Power Supply, AC Power Cord, DC Power
Cord
1. Disconnect cords from the device and power source before cleaning.
2. Clean the cords as needed by using a damp cloth with a mild
household cleaner, and wipe dry.
Cannula and Tubing
Clean and replace the cannula and tubing according to the manufacturer’s
instructions.
Cleaning
Suggested
cleaning
interval
Number of
cleaning
cycles *
Compatible cleaning
method
Outer Cabinet
and Battery
7 days 260 Mild dish soap (2 tbsp)
and warm water (2 cups)
Power Supply
and Cords
7 days 260 Mild dish soap (2 tbsp)
and warm water (2 cups)
Carrying Case 7 days 260 Mild dish soap (2 tbsp)
and warm water (2 cups)
* number of cleaning cycles determined by suggested cleaning interval and
expected service life
Cleaning & Disinfection When There Is A
Change Of Patient
The iGo2 POC should be cleaned and disinfected between patients as
follows:
1. Dispose of all accessories that are not suitable for reuse, particularly
the oxygen tubing and nasal cannula.