ELECTROMAGNETIC COMPATIBILITY INFORMATION
If the device performance is lost or degraded due to electromagnetic disturbances, a visual and audible alarm indicates that the device is not meeting
specication or a failure has been detected.
WARNING
MR Unsafe
Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to
the iGo2 or MR medical devices. The device and accessories have not been evaluated for safety in an MR environment.
Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic
security systems (metal detectors) as it may cause unacceptable risk to the patient or damage to the iGo2. Some electromagnetic sources may not
be apparent, if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power
cord and discontinue use. Contact your home care provider.
This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of
an ME SYSTEM for magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high.
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
Cables and maximum lengths of cables
DC power cord (cigarette lighter adapter) #DV6X-619 maximum length = 2.2 meter (7.2 ft)
WARNING
Use of accessories and cables other than those specied or provided by the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the iGo2, including cables specied by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.