LT-2334
15
TROUBLESHOOTING AND ALARM SYSTEM
Overview of Alarms
This device contains an alarm system which monitors the state of the device and alerts of abnormal operation, loss of essential performance or failures. Alarm
conditions are shown on the LCD display. The alarm system functions are tested at power up by lighting all visual alarm indicators and sounding the audible
alarm (beep). All alarms are Low Priority Technical Alarms.
To mute the Audible Alarm, press and release power button. The Alarm Silenced symbol will appear on the display until the alarm condition is corrected.
Alarm
Condition
Display
Shows
Visual Alarm Signal
Meaning
Audible Alarm
Signal
Visual Alarm Signal
Cleared by Action to take
Battery nears
depletion
Flashing
Battery is <10% Yes, when battery is <
5%
Turn off device or switch to
alternate power source
Charge battery, change battery, or
switch to alternate power source
Low Oxygen
Concentration
Oxygen concentration is
<86%
Yes, if oxygen drops
below 85%
Turn unit off unless the
device is still in startup
period
Call provider and switch to reserve
oxygen.
Start-up period
Device and systems are
tested at power up
No Once oxygen reaches at
least 86%
Wait until startup is nished (up to
10 minutes)
Malfunction
RED Service Required
symbol ashes indicating
malfunction
Yes Turn unit off Call provider and switch to reserve
oxygen.
No Breathing
Detected
None More than 60 seconds
between requested
dispense cycles.
Yes Turn unit off or breathe on
device.
Resume breathing. If no breathing
is detected, device will go into
AutoBreath™ mode and dispense
oxygen at regular intervals until
breathing is detected again.