CLEANING & MAINTENANCE
Mattress
8) Remove any soiled or used bedding.
9) Examine the mattress for visible soilage of bodily uids.
10) If no disinfection is required, brush off or wipe down all surfaces of the cover sheet with soap and
water before wetting with any liquid disinfectant.
11) If disinfection is required, follow the procedure below:
a. Use rubber gloves and eye protection.
b. Unzip the top cover from the mattress.
c. Prepare detergent/disinfectant solution (registered by the EPA recommended) according to
the preparation recommended for correct use-dilution.
d. 1:9 Bleach and water dilution may be used.
e. With the mattress fully deated, wipe down all surfaces around and in between the air cells,
including the cells.
f. Covers may be immersed and soaked in disinfectant for the required incubation period. After
pre-soaking, the cover may be rinsed through a regular cycle in a washer with no soap then
laundered with mild detergent (wash temperature 93°F/34°C, rinse temperature 78°F/26°C or
on the coldest setting).
g. Allow all covers and parts to aerate until they are fully dry.
12) Repeat the process with the tubing set: spray/wipe, incubate, and air dry.
13) Dry the mattress on a at surface area after cleaning, away from exposure to the sun.
14) Avoid long exposure to sunlight.
HANDLING AND STORAGE
• Lay the mattress out at and upside down.
Roll from the foot end towards the head end; the foot-end strap can then be stretched around the rolled
mattress to prevent unrolling.
• Do not fold, crease or stack the mattress.
MAINTENANCE
General
• Check the power cord and plug to see if there are abrasions or excessive wear.
Check the mattress cover for signs of wear or damage. Ensure the mattress cover and tubes are
connected correctly.
Plug in the control unit and check the airow from the hose connection port. The airow should alternate
between ports every half-cycle time.
Check the air hoses to see if there are any kinks or breaks. For replacement, please contact your local
agent or dealer.