Spezifikationen
Function Bereich Auflösung Genauigkeit Anmerkung
% of span
Messen mVolt 0 -120 mVDC 0.01 mV + 0.025% R - input > 20MΩ
Ausgang mVolt -10 - +120 mVDC 0.01 mV + 0.025% R - output 0.2Ω
Messen Volt 0 - 120 VDC 0.01 V + 0.05% R - input > 1MΩ
Ausgang Volt 0 -12 VDC 0.001 V + 0.025% R - output 0.2Ω
Messen mA 0 - 52 mA DC 0.01 mA + 0.05% R - input 10.5Ω abges.
Ausgang mA 0 -24 mA DC 0.01 mA + 0.05% R - max. 900Ω
Transmitter Sim. 0 - 24 mA DC 0.01 mA + 0.05% V - max. ext. 56 Volt
Mess./Sim. Ohm 0 - 390Ω 0.1Ω + 0.1Ω 1.25 mA Ausgangsstrom
Mess./Sim. Pt100 -200/850
0
C0.1
0
C + 0.25
0
C bei Ohm und Wider-
Mess./Sim NI100 - 60/250
0
C0.1
0
C + 0.25
0
C standsthermometer
Mess./Sim NI120 - 80/260
0
C0.1
0
C + 0.25
0
C Messungen
Mess./Sim. Cu 10 -200/260
0
C2
0
C + 2.0
0
C
Mess./Sim. TC-J/L -100/1190
0
C0.2
0
C + 0.4
0
C
Mess. Sim. TC-J/L -210/ -100
0
C0.4
0
C + 1
0
C
Mess./Sim. TC-K -210/1370
0
C0.3
0
C + 0.6
0
C
Mess./Sim. TC-K -230/- 210
0
C1
0
C + 2
0
C
Mess./Sim. TC-T/U -75/400
0
C0.2
0
C + 0.5
0
C
Mess./Sim. TC-T/U -180/- 75
0
C0.5
0
C + 1
0
C
Mess./Sim. TC-T/U -250/-180
0
C0.7
0
C + 1.6
0
C