From lab to production,
providing a window into the process
-2-
www.dynisco.com
Rev: n/aP/N: n/a ECO: n/a
CAUTION: Installaon should be only performed by technically competent
personnel. Local Regulaons regarding electrical installaon & safety must be
observed. The host equipment is required to provide a suitable electrical, mechanical
and re enclosure to meet relevant safety standards. Impairment of protecon will
occur if the product is used in a manner not specied by the manufacturer.
CAUTION: All power supply connecons to the device must be removed when
carrying out any form of maintenance.
1. Installaon
Installing Opon Modules/Maintenance
CPU PCB
Opon Module 2
Opon Module 1
Mounng Struts
Opon Module 3
PSU PCB
Future Opons
To access modules, rst detach the PSU and CPU boards from the front by liing rst the upper, and
then lower mounng struts. Gently separate the boards.
a. Plug the required opon modules into the correct connectors, as shown below.
b. Locate the module tongues in the corresponding slot on the opposite board.
c. Hold the main boards together while relocang back on the mounng struts.
Replace the instrument by aligning the CPU and PSU boards with their guides in the housing, then
slowly push the instrument back into posion.
NOTE: Opon modules are automacally detected at power up.