6.0 TECHNICAL SPECIFICATIONS
6.1 Technical Specifications
Autozero: Automatic autozero cycle.
Dilution (where provided): Expansion system of the CO sensor measuring range up to 100,000ppm
(10.00%) programmable as a simple protection of the CO sensor with
triggering threshold programmable by the user. Preset triggering threshold at
1500 ppm.
Gas measurement sensors: Up to 4 configurable sensors: Electrochemical and Pellistor
Self-diagnosis: All the functions and internal functions are checked and errors signalled.
Temperature measurement: Double K thermocouple input with mini connector (ASTM E 1684-96) to
measure differential temperature (supply and return)
Measurement of ambient temp.: Via internal sensor or T2 thermocouple input with remote probe.
Fuel Types: 12 predefined by the factory and 16 that can be programmed by the user.
Power: Li-Ion battery pack with internal protection circuit.
Battery charger: External 5Vdc 2A battery charger with female A-type USB connector +
connection to the device with the same serial communication cable supplied.
Charging time: 5 hours to charge from 0% to 90% (6 hours for 100%). The device can also
be charged by connecting it to the PC, the device must be turned off, the
charging time depends on the output current from the PC and may be more
than 12 hours.
Instrument working time: 12 hours of non-stop operation (excluding printing).
Printer: Thermal integrated with easy loading paper and sensor for the presence of paper
Printer powered: By the analyzer batteries.
Printer autonomy: Up to 40 analysis reports with the batteries fully charged.
Internal data memory: 2000 complete data analyses, time and name of the customer can be stored.
User data: 8 programmable operator names.
Print-out heading: 4 lines x 24 characters, customizable by the user.
Display: Graphic 272 x 480 pixels, backlit, color TFT 4.3".
Communication port: USB with B-type connector
Bluetooth (where provided): Communication range: <100 meters (free field) (Class 1)
Line filter: With replaceable cartridge, 99% efficient with 20um particles.
Suction pump: 1.0 l/min heads at the flue up to 135mbar.
Measurement of flow: Internal sensor to measure the flow of the pump.
Condensate trap: External
Smoke Pump: Using an external manual hand pump; it is possible to enter and print the
smoke index.
Leak test: Gas pipes tested for leaks with separate printout of the result, by means of
the attachment AACKT02, according to UNI 7129 (new systems) and UNI
11137: 2012 (existing systems), with automatic calculation of pipe volume.
Condensing boiler efficiency: Automatic recognition of the condensing boiler, with calculation and printout
of efficiency (>100%)
Environmental gases: Measurement and separate printout of the ambient CO values.
Draft test: Draft tested as per the UNI 10845 standard. By using the internal sensor
connected to the port P+, resolution 0.1 Pa , accuracy 0.5 Pa.
Operating temperature range: -5°C to +45°C
Storage temperature range: -20°C to +50°C
Operating humidity range: 20% to 80% RH
Protection grade: IP42
Air pressure: Atmospheric
Outer dimensions: Analyzer: 3.5” x 12.2” x 2.4” (9 x 31 x 6 cm) (L x A x P)
Case: 6” x 19” x 15” (15 x 48 x 38 cm) (L x A x P)
Weight: Analyzer: ~ 2 lbs (0.9 Kg)
Compliant with the European standard EN50379-1 and EN50379-2 for the following measurements:
O
2
CO m
edium
NO
Temperature (flue gas)
Temperature (supply air)
Pressure (draft)
Pressure (differential)