48
@IST01530EEJ 019175A0 190411
the following data must be configured:
Memory: use this submenu to define the memory in which the test data and client details are to be stored.
Fuel: the user will be asked to define the type of fuel used by the plant.
Operator: this is where the name of the test operator can be entered.
Mode: by entering this submenu, the user can determine the analysis mode - manual or automatic.
If automatic mode is chosen, the reading duration of each and every test must be set, besides the printing mode
- manual or automatic. When flue gas analysis begins, the instrument will perform and memorize the three tests
automatically, at the respective intervals set (at least 120 sec. according to UNI 10389-1).
At the end of each test the instrument will emit an audible alarm (one beep” after the first test, two beeps” after
the second test and three “beeps” after the third test).
At this point, when all three tests are over, if “Manual Printing has been chosen the instrument will display the
average of the three tests with the possibility of recalling the individual values.
If desired, the user can then print the relative data (total, complete, etc....). On the contrary, if “Automatic
Printingwas selected, the instrument will print the test data automatically, based on the current print settings,
without displaying the average test values.
Caution: w hen in automatic mode Draft, Smoke and ambient CO (NO) measurements must be taken
before initiating the flue gas analysis.
If, on the other hand, manual analysis mode is chosen, flue gas analysis will proceed manually (please see
relative Flow Chart). In this case the print settings and automatic test duration will not be considered.
At this point manual analysis may commence, first waiting at least two minutes until the displayed values
stabilize: The user can then proceed with data storage, if required, or print the analysis report directly.
The latter will be printed in the format set beforehand.
When all three tests are over, the user can recall the average analysis screen containing all the data necessary
for compiling the maintenance log of the boiler or plant.
In both automatic and manual modes, all the pollutant values (such as CO / NO / NO
x
) can be translated into
normalized values (referenced to the previously defined O
2
level) by simply pressing the button .
4.16.4 End of Analysis
At the end of the combustion analysis, carefully remove the sample probe and remote air temperature probe, if
used, from their relative ducts, taking care not to get burnt. Then carefully disconnect the water trap, hoses,
and sampling probe from the analyzer.
Switch off the instrument by pressing the On/Off key.
At this point, if the instrument has detected a high concentration of CO and/or NO, a self-cleaning cycle will be
initiated during which the pump will draw fresh outside air until the gas levels drop below acceptable values.
At the end of the cycle (lasting no longer than 3 min.) the instrument will switch itself off automatically.
1
2
3
4
5
6
7
8

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the E Instruments 4400 and is the answer not in the manual?

E Instruments 4400 Specifications

General IconGeneral
BrandE Instruments
Model4400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals