2.3 Measurement Ranges and Accuracies
All data relative to concentration accuracies are referred to an instrument operating at a constant temperature
within the correct operating range (-5°C .. +45°C), being in operation for at least 15 minutes, powered by its
internal battery and after completion of auto-zero procedure.
Notes: (1) The maximum CO
2
value displayed depends on the type of fuel.
(2) Stated precision includes error of the external sensor RTD Pt100 class A DIN 43760 (1980).
(3) Stated precision includes error of the external sensor type K thermocouple class 1 IEC584.
(4) Pressures greater than 750 hPa may permanently damage sensors or impair their characteristics.
1
2
3
4
5
6
7
8
MEASUREMENT SENSOR RANGE RESOLUTION ACCURACY
O
2
Electrochemic al sensor 0 .. 25.0% vol 0.1% vol ±0.2% vol
CO
with H
2
compensation
Electrochemic al sensor 0 .. 8000 ppm 1 ppm
±10 ppm 0 .. 200 ppm
±10% meas ured value 2001 .. 8000 ppm
diluted
Electrochemic al sensor 0.15 .. 5.00% vol 0.01% vol ±20% meas ured value
CO Low range
with H
2
compen sation
Electrochemic al sensor 0 .. 500 ppm 0.1 ppm
±2 ppm 0 .. 40.0 ppm
±5% measured val ue 40.1 .. 500.0 ppm
diluted
Electrochemic al sensor 100 .. 3125 ppm 10 ppm ±20% meas ured value
CO
Mid range
Electrochemic al sensor 0 .. 20000 ppm 1 ppm
±100 ppm 0 .. 2000 ppm
±10% meas ured value 4001 .. 20000 ppm
diluted
Electrochemic al sensor 0.3 .. 12.5% vol 0.01% vol ±20% meas ured value
CO
Hi range
Electrochemic al sensor 0 .. 10.00% vol 0.01% vol
±0. 1% vol 0 .. 2.00 %
±5% measured val ue 2. 01 .. 10.00 %
NO
Electrochemic al sensor 0 .. 5000 ppm 1 ppm
±5 ppm 0 .. 100 ppm
±5% measured val ue 101 .. 5000 ppm
NO Low range
Electrochemic al sensor 0 .. 500 ppm 0.1 ppm
±2 ppm 0 .. 40.0 ppm
±5% measured val ue 40.1 .. 500.0 ppm
NOx
Calculated
SO
2
Electrochemic al sensor 0 .. 5000 ppm 1 ppm
±5 ppm 0 .. 100 ppm
±5% measured val ue 101 .. 5000 ppm
SO
2
Lo w rang e
Electrochemic al sensor 0 .. 500 ppm 0.1 ppm
±2 ppm 0 .. 40.0 ppm
±5% measured val ue 40.1 .. 500.0 ppm
NO
2
Electrochemic al sensor 0 .. 1000 ppm 1 ppm
±5 ppm 0 .. 100 ppm
±5% measured val ue 101 .. 1000 ppm
NO
2
Low range
Electrochemic al sensor 0 .. 500 ppm 0.1 ppm
±2 ppm 0 .. 40.0 ppm
±5% measured val ue 40.1 .. 500.0 ppm
CxHy
Pellistor sensor 0 .. 5.00% vol 0.01% vol ±0.25% vol
CO
2
Calculated 0 .. 99.9% vol 0.1% vol
CO
2
NDIR sensor 0 .. 40.0% vol 0.1% vol
±0. 3% vol 0.0 .. 10.0 %
±5% measured val ue 10.1 .. 40.0 %
Air temperatur e
Pt100 s ens or -20.0 .. 120.0 °C 0.1 °C ±0.5 °C
Flue gas temper ature
TcK sens or -100.0 .. 1250.0 °C 0.1 °C
±0. 5 °C 0 .. 100 °C
±0. 5% meas ured value 101 .. 1250 °C
Pressure
(draught & differential)
Piezoel ectric sensor -10.00 .. 200. 00 hPa 0.01 hPa
±1% measured val ue -10.00 . . -2. 01 hPa
±2 Pa -200 . . 200 Pa
±1% measured val ue 2.01 .. 200.00 hPa
Differential temp erature
Calculated 0 .. 1250.0 °C 0.1 °C
Air index
Calculated 0.00 .. 9. 50 0.01
Stack loss
Calculated 0.0 .. 100.0 % 0.1 %
Efficiency
Calculated 0.0 .. 100.0 % 0.1 %
Efficiency
(condensing)
Calculated 0.0 .. 120.0 % 0.1 %
Smoke index
External instrument 0 .. 9
Excess ai r
Calculated 0 .. 850 % 1 %
(1
(2
(4
(3