4.2 Weight Scale - Error Codes
Overloaded condition.
Wrong connection at load cell.
Load cell has quality problem.
Reduce the weight.
Check load cell connection.
Inspect load cell for damage.
Check the input and output.
Bad connections.
Load cell has quality problem.
Check scale is resisted or not, foot is kept level
or not.
Check load cell connection.
Check input and output resistance of load cell.
Overweight calibration error.
Incorrect weight input or incorrect
weight on weighing platform.
Input weight correctly during calibration.
Place correct weight on weighing platform.
Underweight calibration error.
Incorrect weight input or incorrect
weight on weighing platform.
The calibration weights Minimum is 10% of
Max. cap. Recommend 60%-80% of Max.
During calibration, the input signal is
negative.
Check all connections. Check load cell. If
connections and load cell are okay, PCB needs
During calibration, the signal is
unstable.
After the platform is stable, start calibration.