EasyManua.ls Logo

EasyTouch ET-201 - Using the Check Strip to Check the Meter

EasyTouch ET-201
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
CAUTION:
●Make sure the “+” and “-“ ends of the batteries match the marks in the battery
compartment.
●Always replace both batteries at same time. Both batteries should be the same brand.
Using the Check Strip to Check the Meter
1. Insert the check strip into the test strip slot on the meter.
2. The screen should display ”OK”. If “X” appears on the screen, remove the check strip
and reinsert it. If “X” displays again, stop all testing and call our customer support.
You need to use the check strip to check the meter when
●You use EasyTouch
®
GCU monitoring system for the rst time.
●You drop the meter.
●You think there is something wrong with the results you got and
you want to check whether meter and test strips are working
correctly or not.
EasyTouch
®
GCU

Table of Contents