18
CAUTION:
Always check the uric acid code key number to make sure it matches the number
labeled on the test strip vial when you run a uric acid control test. Otherwise you may
get incorrect results.
Always write down the opening date of your uric acid control solution vial and test strip
vial. Both of them are good for only 2 months after rst opening or until the expiration
date, whichever comes rst.
● The result will be stored in memory and appear just like a blood uric acid result.
1. Insert the uric acid code key into the code key
slot on the back of the meter.
2. Take one strip from the vial. Close the vial
quickly.
3. Insert the test strip into the test strip slot on
the meter. The meter will rst display the code
number, and then the blood symbol “
S
”.
Steps for Running a Uric Acid Control Check
Caution:
Make sure that the code number shown on the screen is the same as the code number
printed on the uric acid test strip vial, otherwise you may get incorrect measurement.