P.39
PM10U
EBARA
CORPORATION
6.3.8 Setting the rotational speed in the SAVE ENERGY mode (S.ENERGY)
Use the up and down arrow keys to change the setting value.
△ : Increase the setting speed by 0.1 kmin
-1
.
▽ : Decrease the setting speed by 0.1 kmin
-1
Upper limit MP, BP : The value lower than the set value for the rated speed
Lower limit MP : 1.0 kmin
-1
, BP : 1.0 kmin
-1
6.3.9 Setting the WARNING value for the back pressure (option)
Use the up and down arrow keys to change the setting value.
△: Increase the setting value by 0.5 kPa.
▽ : Decrease the setting value by 0.5 kPa.
Upper limit : 30.0 kPa
Lower limit : 5.0 kPa
Factory setting : 20.0 kPa
WARNING reset value : Set value –2.0 kPa
Use the keys, ”△” and ”▽” to set.
SET ALARM SP
BACK PRES.?
PRG.
ENTER
ENTER
B