38
8.6.2 Extra Cycle
Program the 30 available steps with
the function and the time in seconds.
8.6.3 Settings Menu
Cycles Ratio: Program the ratio of the ANALYSIS CYCLES versus the EXTRA CYCLES (for example ANALYSIS 20 &
EXTRA CYCLE 1 means that every 20 ANALYSIS CYCLEs the analyzer will perform 1 EXTRA CYCLE.
Result alarm: Program the low / high
alarm value. The alarm is expressed in
percent of the full scale.
Output signal: Allows the user to enable
or disable the 4-20 mA output.
Sample alarm: Allows the user to enable
or disable the Loss of Sample Alarm.
CYCLE WAIT Enter the WAIT Time for
Step 30 of the ANALYSIS CYCLE
8.6.2 Calibration Menu
Allows the user to enter the value of the standard used for calibration
Perform manual Blank and Span
Calibration (refer to Section 7)
Displays:
Calibration Standard Used
Calibration Factor
Blank absorbance value
Calibration Date and Time
Last Calibration Date and Time
and Factor