SD Series Dry Cabinets Setup and Operation Manual | A55-7379-00 Rev: 2.0.0 Page 56
10.2 ECD SMARTDRY™ HUMIDITY AND TEMPERATURE SENSOR
This sensor is a highly accurate environmental sensors. Exposing the sensor to the following can cause drift of
the humidity reading irreversibly in most cases. The list below are examples and do not represent a com-
plete list of possible harmful substances, but are the most common used in electronic manufacturing.
Do not allow the sensor to be exposed to:
Solvents or other organic compounds including: Ketenes, Acetone, Ethanol, Isopropyl Alcohol, Tolu-
ene, etc.
Acids and bases including: HCl, H2SO4, HNO3, NH3 etc.
Ozone in high concentration or H2O2
Cleaning agents or strong air blasts from an air-gun.
PLEASE NOTE THAT SUCH CHEMICALS ARE CONTAINED IN UNCURED EPOXIES, GLUES, ADHE-
SIVES, ETC. AND OUTGAS DURING BAKING AND CURING. DO NOT USE THE SMARTDRY™ CABINET
AS A CURING CHAMBER FOR THESE PROCESSES. THESE CHEMICALS ARE ALSO ADDED AS PLASTI-
CIZERS IN PLASTICS, USED FOR PACKAGING MATERIALS, AND MAY ALSO OUT GAS FOR SOME
TIME. ONLY ALLOW WELL CURED APPROVED MOISTURE BARRIER BAG TO BE USE IN THE
SMARTDRY™ CABINET.