EasyManuals Logo

ECD SmartDRY SD-30 User Manual

Default Icon
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
SD Series Dry Cabinets Setup and Operation Manual | A55-7379-00 Rev: 2.0.0 Page 75
11.4 SYSTEM ALERTS
Door Open - Cabinet door has been open for more than 30 seconds.:
o Close door(s).
o Door switch may have become broken. Contact Support
o Door switch actuator is missing/improperly adjusted.
Dryer Failure - Dryer has been regenerated twice and has failed to reduce the humidity to within 5%
RH of the set point.:
o Confirm dryer cable has not become loose or disconnected.
o Replace the dryer. (Refer to topic 10.4)
RH Probe Read Failure: The controller has failed to communicate RH probe:
o RH Probe may have become damaged or bad. (Refer to topic 10.5)
o Confirm cable has not become loose or disconnected at probe or Controller Main Board. (Refer to
topic 10.5).
SmartBAKE™ Heater Failure: The baking heater has failed to reach temperature set point within 75
minutes:
o Confirm cable has not become loose or disconnected at Power Board. (Refer to topic 10.6)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ECD SmartDRY SD-30 and is the answer not in the manual?

ECD SmartDRY SD-30 Specifications

General IconGeneral
BrandECD
ModelSmartDRY SD-30
CategoryDryer
LanguageEnglish

Related product manuals