7 TROUBLESHOOTING
Observation
Action
Sample will not
draw into Settling
Cylinder
- Check the air line isolator is open (part of the filter/regulator
unit).
- Check air pressure on regulator. Air pressure should be regulated
to 6 bar. If there is insufficient air pressure, the vacuum may not
be strong enough to draw the sample.
- Check airline push-fit connections are firmly screwed into top
cap of Settling Cylinder.
- Check air hoses are completely pushed into push-fit connections.
- Check Settling Cylinder for air leaks. Ensure top union and
instrument unions are tight. Check that union O-rings are
installed.
- Check turbidity instrument connection is airtight. Apply thread
tape to instrument and re-install if connection feels loose.
- Check sample line camlock is connected firmly and square.
- Check camlock female connection O-ring. If O-ring is worn or
missing, replace.
- Listen for air leaks
Sample level will
not hold in cylinder
- Excluding the Settling Cylinder top union, air leaks will be evident
with air bubbles in the cylinder.
- Check Settling Cylinder for air leaks. Ensure top union and
instrument unions are tight. Check that union O-rings are
installed.
- Check turbidity instrument connection is airtight. Apply thread
tape to instrument and re-install if connection feels loose.
- Check sample line camlock is connected firmly and square.
- Check camlock female connection O-ring. If O-ring is worn or
missing, replace.
Settle timer does
not record
- Flocculant not available to system pump failure, Flocculant
makeup issues, line blockage. Check Flocculant system operation,
dosing pump operation and test flocculant solution samples.
- Turbidity instrument fouled. Remove instrument and clean.