EasyManuals Logo

EDAN CADENCE II User Manual

EDAN CADENCE II
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
CADENCE II Fetal Monitor User Manual
- 3 -
A SHOCK HAZARD - The power receptacle must be a three-wire grounded outlet. A
hospital grade outlet is required. Never adapt the three-prong plug from the monitor
to fit a two-slot outlet. If the outlet has only two slots, make sure that it is replaced
with a three-slot grounded outlet before attempting to operate the monitor.
A Do not apply this monitor and other ultrasonic equipment simultaneously on a same
patient, in case of possible hazard caused by leakage current superposition.
A Do not apply this monitor simultaneously with other PATIENT-connected equipment,
such as, a cardiac pacemaker or other electrical stimulators, on the same patient.
A Do not switch on device until all cables have been properly connected and verified.
A Do not touch the signal input or output connector and the patient simultaneously.
A Equipment and devices that connect to the monitor should form an equipotential
body to ensure effective grounding.
A Disconnect the power cord before changing fuses. Replace them with those of the
same specifications only.
A SHOCK HAZARD - Do not attempt to connect or disconnect a power cord with wet
hands. Make certain that your hands are clean and dry before touching a power
cord.
A SHOCK HAZARD - Do not remove the top panel cover during operation or while
power is connected.
A The monitor is not protected against defibrillation. Do not apply it during
electro-surgery or MRI; otherwise it might result in harming the patient or the
operator.
A To ensure the proper working of the monitor, only connect accessories supplied or
recommended by the manufacturer to the device.
A Accessory equipment connected to the analog and digital interfaces must be
validated according to the respective IEC/EN standards (e.g. IEC/EN 60950 for
data processing equipment and IEC/EN 60601-1 for medical equipment).
Furthermore all configurations shall comply with the valid version of the system
standard IEC/EN 60601-1-1. Everybody who connects additional equipment to the
signal input connector or signal output connector configures a medical system, and
is therefore responsible that the system complies with the requirements of the valid
version of the system standard IEC/EN 60601-1-1. If in doubt, consult our technical
service department or your local distributor.
For proper monitoring:
A This device is not intended for treatment.
A The fetal spiral electrode and intrauterine pressure catheter are disposable. Discard

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN CADENCE II and is the answer not in the manual?

EDAN CADENCE II Specifications

General IconGeneral
BrandEDAN
ModelCADENCE II
CategoryMonitor
LanguageEnglish

Related product manuals