iSE Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 23 -
Chapter 5 Inspection and Troubleshooting
5.1 Appearance Inspection
Perform the following inspection procedures before operation.
Check if there are any obvious defects on the electrocardiograph and patient cable.
Check if there are any obvious defects on the external cables, power plug and power
cable.
Ensure that all the exposed screws are firmly fixed.
Check if there are any crackles, rifts or distortions on the outer-casing of the device.
If any failure is detected, use the ECG Device only after maintenance has been carried out by the
service personnel of the manufacturer or the distributor.
5.2 Power-On Test
Connect the electrocardiograph with the mains supply, check if the power light turns orange.
Switch on the electrocardiograph and carry out the following inspections:
The LCD screen is lit and EDAN logo is displayed.
You can hear the start-up sound. The iSE screen is displayed and can be operated by
tapping.
If any failure is detected, refer to section 5.8 "Troubleshooting" for details.
5.3 Checking Battery Charging
1. Insert the batteries into the battery compartment. Connect the electrocardiograph to the AC
mains; check if the power light turns orange.
2. Switch on the electrocardiograph and wait until the main screen appears. Check whether the
battery charging symbol is shown at the top right corner of the screen. When the battery is
full, the power light will turn green.
3. Disconnect the electrocardiograph from the AC mains. The power light turns blue. Let the