SE-3&SE-300 Series Electrocardiograph User Manual Introduction
- 13 -
Managing ECG Records.
4) 1mV/COPY Key
(SE-3) (SE-300 series)
In the Manual mode, press this key to print a 1mV calibration mark while printing ECG
reports.
In the AUTO or RHYT mode, press this key to copy and print the recent report.
5) MODE/RST Key
(SE-3) (SE-300 series)
Press this key to select a printing mode among AUTO, MANU, OFF-AUTO, RHYT,
OFF- RHYT and R-R.
The switch order of lead groups in each mode is listed in Table 2-1.
In the MANUAL mode, press this key to reset the waveform quickly while printing ECG
reports.
WARNING
When using the device with a defibrillator, after the defibrillator discharges, you should
press the MODE/RST key to reset the waveform quickly.
NOTE: The detailed information of the automatic mode can be set on the System Setup
Screen.
Table 21 Switch Order of Lead Groups in Different Modes
Mode
AUTO(Standard)
І/II/III
aVR/aVL/aVF
V1/V2/V3
V4/V5/V6
AUTO(Cabrera)
aVL/ I /-aVR
II /aVF/ III
V1/V2/V3
V4/V5/V6
MANUAL