A419-00-880 Issue F
Page 34 © Edwards Limited 2007. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Installation
3.11 Connect to your control equipment
CAUTION
Do not connect voltages greater than 30 V to the EPX interface control. If you do, the interface control will not
comply with the low voltage safety directive and the equipment may be damaged.
The pump may be connected to your control equipment through the Tool Interface connector (TIM). The TIM is
supplied with a protective plastic cap fitted. Remove this before connecting the mating connector. Several different
tool interfaces are available. The label next to the tool connector identifies which TIM is fitted, refer to Table 12 to
identify the tool connectors and mating tool connector kits available (not supplied with the pump).
Note: Plug/Socket designation refers to the pins of the connector.
3.11.1 Inputs
Refer to Table 13 for a definition of the input signals to the pump system.
Notes: 1. Volt free contacts sensed at 24 V / 5 mA, these inputs should not be linked together.
2. Voltage inputs accept AC or DC voltage (except SPI pins 15 and 16), between 15 V and 24 V selects the
input as active, below 5 V selects inactive.
3. To ensure correct pump operation, at least 10 seconds must be allowed between successive operations
of the input signals.
WARNING
Tool interface control signals are for control purposes only and should not be relied upon for safety
critical functions.
Table 12 - Tool connectors and mating tool connector kits
Identification Part Number Tool Side Connector Tool Connector Kit
SPI D373-60-310 CPC 16/17 Socket D374-20-801
MCM D373-60-320 CPC 16/17 Socket D374-22-802
TEL D373-60-330 25w D Type Socket D374-20-802
E73 D373-60-340 15w D Type Plug D374-20-803
LAM D373-60-350 25w D Type Plug D374-22-801
C3 D373-60-360 9w D Type Socket D374-21-803
HIT D373-60-370 37w D Type Socket D374-21-804
Table 13 - Tool input signals to the pump system
Function Signal Type Pins Operation
Pump On/Off Voltage input SPI - 1 and 2
MCM - 1 and 2
LAM - 10 and 23
C3 - 2 and 6
Pump runs when voltage
applied
Volt-free contact input MCM - 5 and 6
TEL - 1 and 14
E73 - 1 and 9
HIT - 7 and 26
(or HIT - 1 and 20)
Pump runs when contact
closed