Observe the storage temperature limits stated in Section 2. Storage below -30 °C (-22 °F) will permanently
damage the pump seals.
Use the following procedure to store the pump:
z Purge the vacuum system and the pump with dry nitrogen and disconnect the pump from the vacuum system.
z Shut down the pump as described in Section 4.
z Disconnect the pump from the electrical supply.
z Disassemble, clean and reassemble the pump. If the pump is to be stored for long periods, loosen the drive
belts and change the oil.
z Place and secure protective covers on the inlet and exhaust ports.
z Store the pump in cool, dry conditions until it is required for use. When required, prepare and install the
pump as described in Section 3. If the pump will be subjected to temperatures below freezing, drain or add
anti-freeze to the water jacket. If the pump will be stored for more than a year, the pump must be cleaned
and overhauled before re-installation.
6.2 Disposal
Dispose of the pump or any components removed from it safely in accordance with local and national safety and
environmental requirements.
Particular care should be taken with components and waste oil which have been contaminated with dangerous
process substances.
Do not incinerate fluoroelastomer seals and O-rings.