WARNING 
 
The surfaces of the STP pump and its peripheral equipment will become extremely 
hot when performing baking. NEVER touch them with bare hands. 
CAUTION 
 
Wind the baking heater around the surface of the STP pump tightly. If the baking heater is not 
wound tightly, the loose parts will overheat. 
 
The baking heater is not included in the scope of supply. 
When preparing the baking heater at your company, refer to the external appearance on each 
Instruction Manual of the STP pump before purchasing it. 
Install the baking heater near the inlet port flange in the perimeter of the envelope. (see Figure 17) 
 
2.3.9  Installation of the air cooling unit 
Refer to "Air Cooling Method" in the STP Pump Instruction Manual (B) for the installation methods 
and other details. 
Note:  The air cooling unit is not available depending on the STP pump model. Refer to the STP 
Pump Instruction Manual (B) and confirm whether the air cooling unit can be used for the 
STP pump. 
 
2.3.10  Installation of the TMS unit 
The TMS unit is an optional accessory. Refer to the STP Pump Instruction Manual (B) for installation 
of the TMS unit. 
Note:  The TMS unit is not available depending on the STP pump model. Refer to the STP Pump 
Instruction Manual (B) and confirm whether the TMS unit can be used for the STP pump.