5  SAFETY FUNCTIONS 
The STP pump is provided with safety functions for various abnormalities/errors. If an 
abnormality/error is found when using the STP pump, check it and take measures in accordance with 
the following procedures. If you cannot trace it or if the STP pump does not function normally after 
the troubleshooting in the STP Control Unit Instruction Manual (C), contact Service office. 
 
5.1  Safety functions 
5.1.1  Power failure 
When the power voltage drops due to a power failure or another fault, the normal operation of the 
magnetic bearing is maintained at the high rotational speed using the regenerative energy of the 
rotating rotor (backup operation during a power failure). 
The lowest rotational speed to which the magnetic bearing can operate at a power failure is called a 
backup rotational speed. The lowest backup rotational speeds vary according to the model of STP 
pump used, refer to "TECHNICAL DATA" in the STP Pump Instruction Manual (B). 
1.  When the rotational speed is at backup rotational speed or more at a power failure: 
• The  STP  control  unit  detects  any  power  failure  of  2  seconds  or  more  and  the  STP  pump 
decelerates. At this time, the "POWER" LED extinguishes and the "FAILURE" LED illuminates. 
• The LCD displays "Mains Failure" or "POWER FAILURE" (SCU-350). An alarm signal is output 
from REMOTE X7 pins (14)-(33) and (15)-(33). (SCU-350 series: I/O TB1 ALARM OUT 
terminals [9]-[10] and [11]-[10]). When the rotational speed reaches the backup rotational 
speed, the rotor lands on the touch down bearing and stops. 
• The "FAILURE" LED extinguishes and the LCD is turned off. The alarm signal output is reset. 
• The STP control unit does not detect a power failure of less than 2 seconds and the STP pump 
will continue to rotate. 
2.  When the rotational speed is less than backup rotational speed at a power failure: 
• The STP control  unit does  not  detect  the  power failure.  The  rotor  lands  on  the touch down 
bearing and stops. Refer to "TECHNICAL DATA" in the STP Pump Instruction Manual (B). 
 
5.1.2  Operation after a power recovery 
The magnetic bearing operation, LED and LCD messages are recovered after a power recovery.  
The "FAILURE" LED extinguishes. 
Press the "START" switch to reaccelerate the STP pump. 
Note:  Establish a procedure so that the power can be supplied to the STP control unit immediately 
after a power recovery.