Inlet connecon NW40
Outlet connecon NW25
Degree of protecon (IEC60529) IP44
3.4 Electrical data
Pumps are supplied set to 200 230 V.
Table 5
Electrical data
Supply (V)
Phase Frequency (Hz) Current (A)
200 230 Single 50 - 60 6.6
100 120 Single 50 - 60 10.6
Table 6
Recommended regional supply protecon
Area
Voltage Rang
UK 230 V 13 A
Europe 230 V 16 A
US 120 V 15 A
Japan 100 V 15 A
3.4.1 Electrical cables
Recommended cordsets and fuses for regional requirements.
Table 7
Recommended cordsets
Descripon Rang Coupler type Item number
Cord set
assembly, UK
Cable Style = H05VVF, 3 x 1.5 mm
2
,
300 V, 70 °C, maximum length of
2.5 metres
Straight entry A50505003
Plug Type = BS1363 UK plug
Appliance coupler = IEC60320 style
C19
Right angled
entry
A50505006
Fuse Type = BS1362 13 amp fuse, to
an IEC60320 style
Cord set
assembly, Europe
Cable Style = H05VVF, 3 x 1.5 mm
2
,
300 V, 70 °C, maximum length of
2.5 metres
Straight entry A50506003
Plug Type = European Schuko VDE
approved, 16 A 250 V rated with
dual earthing contact
Appliance coupler = IEC60320
style C19
Right angled
entry
A50506006
Cord set
assembly,
Cable style = SJT, 3 x 14 AWG, 300 V,
90 °C, VW1 maximum length of
3 metres
N/A
Page 13
A73101880_F - Technical data

Table of Contents

Related product manuals