Do not touch or inhale the thermal breakdown products of uorinated materials
which may be present in the pump if the pump has been heated to 260 °C and
above. Fluorinated materials are safe in normal use but can decompose into very
dangerous substances (which may include hydrouoric acid) if they are heated to
260 °C and above. The pump may have overheated if misused or in a re. Safety
Data Sheets for uorinated materials used in the pump are available on request,
contact us or the supplier.
6.2 Maintenance plan
The plan shown in Table: Maintenance plan on page 26 details the roune maintenance
operaons necessary to maintain the pump in normal use. Instrucons for each
operaon are given in the secon shown.
More frequent maintenance may be required if the pump is used to pump corrosive or
abrasive gases and vapours.
Table 9
Maintenance plan
Operaon Frequency Refer to Secon
Inspect and clean inlet
strainer
Annually
Inspect and clean the inlet
strainer on page 26
Inspect and clean the gas
ballast control (if ed)
Annually
Inspect and clean the gas
ballast control (if ed) on page
27
Clean the external fan cover Annually
Clean the external fan cover on
page 28
Replace the p seal and
exhaust valve
Annually or as required Replace the p seals on page 28
Test the motor/drive
condion
15000 hours
Test the motor condion on
page 28
Replace the bearings 35000 hours Contact us
6.3 Inspect and clean the inlet strainer
Whenever disconnecng the pump from the vacuum system:
Remove any debris trapped by the inlet strainer (in the inlet port).
Inspect the inlet strainer and if necessary, clean with a cleaning soluon suitable
for the substances pumped. Ret the inlet strainer before reconnecng the pump
to the vacuum system.
1.
Refer to Figure: Inlet strainer assembly on page 27. Disconnect the vacuum system
from the pump inlet port (item 3) and remove the centring ring and strainer
assembly (item 1) and the Oring (item 2). Inspect the centring ring and the Oring.
If clean, connue at Step 5. If not clean, connue at Step 2.
2.
Remove the Oring (item 2) from the centring ring and strainer assembly (item 1).
Do not allow the Oring to come into contact with the cleaning soluon.
3.
Wash the centring ring and strainer assembly in a suitable cleaning soluon and
allow it to dry.
4.
If necessary, wipe the Oring with a clean, dry, lintfree cloth.
Page 26
A73101880_F - Maintenance