5.7 Shut down
WARNING:
When the pump has been switched o the fan will connue to run for 1 minute.
Use the procedure below to shut down the pump:
1.
If shung the pump down prior to a period of storage, remove any process gases
by running on high ow gas ballast for at least one hour.
2.
Close any vacuum system isolaon valve to prevent suck‑back into the vacuum
system (where ed).
3.
Switch o the pump. (When the pump is switched o the fan will connue to run
for 1 minute).
4.
Vent the pump by the gas ballast control (if ed) or valve on the inlet.
Figure 4 Logic interface schemac
1. Speed control pins 3, 4 and 5
Maximum current per pin = 7 mA when
linked to pin 1 (24 V)
2. Pump healthy signal
Signal high = normal (maximum
permissible current = 100 mA)
1. Speed control pins 3, 4 and 5
Maximum current per pin = 7 mA when
linked to pin 1 (24 V)
2. Pump healthy signal
Signal high = normal (maximum
permissible current = 100 mA)
Table 8
Pin status on the logic interface connector
Mode
Speed
percentage
Pin 1 Pin 3 Pin 4 Pin 5
Normal 100% 24 V Link open open
Boost 116% 24 V Link Link Link
Idle 67% 24 V Link Link open
The tolerance of the power supply can be ± 10%. Make sure all the unused pins are not
connected.
Page 24
A73101880_F - Operaon