Leak test the system aer installaon is complete and seal any leaks found to
prevent leakage of hazardous substances out of the system and leakage of air into
the system.
Mechanical liing equipment should be aached to the liing eye on the pump.
Loose slings should not be used to li the pump.
4.2 System design consideraons
WARNING:
Ensure the exhaust pipeline cannot become blocked. If an exhaust isolaon valve is
ed, ensure the pump cannot be operated with the valve closed.
Consider the following points when designing the pumping system:
We recommend the use of a foreline vacuum isolaon valve to allow the pump to warm
up before pumping condensible vapours or if the vacuum needs to be maintained when
the pump is not running.
Avoid high levels of heat input into the pump from the process gases, otherwise the
pump may overheat and cause the thermal overload device to open.
If the pump is used in a high ambient temperature with high gas throughput, the pump
body temperature may approach 65 °C. We recommend the use of addional guarding
to prevent contact with hot surfaces under these condions.
Dilute ammable mixtures to safe concentraons by providing an inert gas diluon
purge. Contact our applicaons team for further advice on diluon requirements if
required.