Test the earth connuity and the insulaon resistance of the pump motor, in accordance
with local regulaons for the periodic tesng of electrical equipment.
We recommend that the earth connuity is less than 0.1 W and the insulaon resistance
is greater than 2 MW. (Reference EN61010‑1).
If the pump fails these tests, contact us.