EasyManuals Logo

EESiFlo EASZ 2Sd-L User Manual

Default Icon
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
EASZ-2 Manual v1.12 - 12.2.2018 Page 25 / 30
7.5. Two-Point Calibration Procedure
This is the recommend form of on-site calibration to ensure that your units are set to a higher accuracy.
If this calibration is performed correctly the accuracy of the readings is ± 0.1%.
Should a higher accuracy of ± 0.05% be required it is necessary to use the factory calibration service from
an EESIFLO authorized service centre.
To calibrate the unit correctly follow the instructions:
1) Ensure that the pipeline is in operation and flowing full of
clean oil of the type and grade required without any air
bubbles or foreign particles.
2) Connect a computer to the EASZ-2 with an original
USB-RS or USB-CL Com Cable.
3) Start the EASZ-2 software and click the “Measuring” tab.
4) Take a sample of the oil from the sampling point just in front
of the EASZ-2 unit. This needs to be sent to the lab for
testing to determine the real water content (%).
5) For each sample of oil taken from the pipeline record the
actual ADC and Temperature values displayed at the top
of the “Measuring” tab.
6) Enter the values into a table and repeat the process for a
wide range of water contents in the oil.
Sample No.
1
2
3
4
5
6
7
8
9
10
ADC Value
20470
20398
20650
20614
20506
20380
20416
22000
20560
20596
Temperature ºC
21.84
21.8
21.68
21.89
21.71
21.81
21.78
22.00
21.56
21.78
Lab Result % Water
0.15%
0.11%
0.25%
0.23%
0.17%
0.10%
0.12%
1.00%
0.20%
0.22%
Max / Min
MIN
MAX
7) After the results of the samples are sent back from the lab
fill in the rest of the table and determine the maximum and
minimum % values (e.g., max is sample 8 and min is 5).
8) Before setting the low and high points in the software
ensure that there is at least a 10 value difference between
the maximum and minimum values.
9) Connect a computer to the EASZ-2 using an original
USB-RS or USB-CL Com Cable.
10) Start the EASZ-2 software, click the “Calibration” tab and
select the Two-Point Calibration option.
11) Add the min ADC and Temp values from the table and the
actual % Water Content value from the lab to the low-point
fields and then press the Set Low-Point button.
12) Add the max ADC and Temp values from the table and the
actual % Water Content value from the lab to the high-point
fields and then press the Set High-Point button.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EESiFlo EASZ 2Sd-L and is the answer not in the manual?

EESiFlo EASZ 2Sd-L Specifications

General IconGeneral
BrandEESiFlo
ModelEASZ 2Sd-L
CategoryMonitor
LanguageEnglish