How-To: Using the IQ-501 with the Fiery server
22
This window will display after all servers in the
list have been calibrated.
21. The Calibration measurement results for both
servers are shown.
Ensure that the D Max values displayed are
within the expected tolerance for the engines
and media selected. Reprint or take appropriate
service steps if not.
22. Click Next to continue with the printer proling
process.
23. Select IQ-501 from the Instrument pull-down.
24. Select an appropriate number of patches from
the Patch set pull-down.
Since these patches will be read automatically
select an appropriate number of patches for the
desired precision or standard. More patches
generally provides better precision.
Note: more sheets will be also be used.
25. Select an appropriate size from the Chart size
pull-down.
26. Click Print to print and measure the proler
patch pages for the rst server.