Guidance and manufacturer’s declaration electromagnetic
immunity
The AirPex is intended for use in the electromagnetic environment
specified below. The customer or the user of the AirPex should assure
that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 8 kV
contact
+/- 2 kV, +/-
4 kV, +/- 8
kV, +/- 15
kV air
+/- 8 kV
contact
+/- 2 kV, +/-
4 kV, +/- 8
kV, +/- 15 kV
air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30 %.
Electrical fast
transients/bursts
IEC 61000-4-4
±2kV
100kHz
repetition
frequency
±2kV
100kHz
repetition
frequency
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Surge
IEC 61000-4-5
Line to line:
±0.5kV,
±1kV
Line to
earth:
±0.5kV,
±1kV, ±2kV
Line to line:
±0.5kV, ±1kV
Line to earth:
±0.5kV,
±1kV, ±2kV
Mains power quality
should be that of a
typical commercial
or hospital
environment.