11 EMC Tables
Page 20 / 25
11.EMC Tables
Guidance and manufacturer’s declaration – electromagnetic emissions
The Fast-Pack is intended for use in the electromagnetic environment specified below.
The customer or the user of the Fast-Pack should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR
11
Group 1
The Fast-Pack uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR
11
Class B
The Fast-Pack is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Not
applicable
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Not
applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The Fast-Pack is intended for use in the electromagnetic environment specified below.
The customer or the user of the Fast-Pack should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+2, 4, 6 kV
contact
+2, 4, 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30 %.