3121-9001-0167_EN.03.00_QT_Manual 04/2016
Running a Quality Control Sample
The Quo-Test A1C Control Kit must be used with Quo-Test A1C Test Cartridges to
ensure that the system is operating correctly.
Only Quo-Test A1C Controls should be used with the Quo-Test A1C Test.
PleasereadtheInstructionsForUseincludedwitheachQuo-TestA1CControlKitto
ensure that the correct procedure is used.
Quality controls should be run:
•WitheachnewlotofTestCartridges.
•WitheachnewshipmentofTestCartridges.
•Atanytimethereisconcernthattheresultmaybewrong,theTestCartridgesmay
not have been stored correctly, the user is unfamiliar with the test procedure or the
user may be performing the test incorrectly.
Use the Quo-Test A1C Control Kit according to the quality standards set by the
organisation or laboratory and according to local and national regulations.
If the controls do not give results within the range stated in the Instructions For Use,
contact your local distributor or customer support.
The quality control result is stored in the memory of the analyzer. The lot number of
theQuo-TestA1CControlKitusedisstoredunderthePatientIDeld.