8 CALIBRATION METHODS (continued)
CALIBRATION METHOD
MODEL
DESCRIPTION
ISO
S, T
Sets the calibration method to Zero Offset and sets the
Counted Average to 5 - in accordance with ISO19840
SSPC PA2
S, T
Sets the calibration method to Rough/2 Point and sets the
Counted Average to 3 - in accordance with SSPC PA2
Swedish
S, T
Sets the calibration method to Rough/2 Point and sets the
Counted Average to 5 - in accordance with SS standards
Australian
S, T
Sets the calibration method to Zero Offset, and sets the
Counted Average to 5 - in accordance with AS standards