Steps:
1. WARNING: First, unplug the generator.
2. Remove the fuse drawer by releasing the two side snaps with a ball-point pen
and pull drawer straight out.
3. Remove the unit’s two fuses.
4. Replace both fuses with supplied 5mm x 20mm 5 AMP, 250 VAC Fast Acting.
Contact Electromed or your service provider for additional fuses.
NOTE: Bad fuses appear identical to good fuses. Therefore, it is important to
discard the old fuses immediately to avoid confusion.